| Lugar de origen: | PORCELANA |
| Nombre de la marca: | AI |
| Certificación: | ISO |
| Número de modelo: | PC4406 |
| Cantidad de orden mínima: | 1 set |
|---|---|
| Precio: | Negociable |
| Detalles de empaquetado: | Panel duro |
| Tiempo de entrega: | 20 días |
| Condiciones de pago: | L/C, T/T, Western Union |
| Capacidad de la fuente: | 20 juegos por mes |
| Voltaje de suministro: | AC220V/50Hz | Estándares aplicables: | GJB-420B, NAS1638, ISO4406 y GOCT17216 |
|---|---|---|---|
| Fuente de luz: | láser semiconductor | rango de tamaño de partícula: | 0.8um~500um |
| Dimensión: | 340*340*640mm | Peso: | Acerca de 28 kg |
| Keyward: | Contador de partículas, probador de limpieza. | Garantía: | 1 año |
![]()
◆The principle of light resistance (shading) method developed by the International Hydraulic Standards Committee;
◆High-precision laser sensor with wide test range, stable performance, low noise and high resolution;
◆Sampling mode of high-pressure injection pump, the sampling volume can be set by itself, the injection speed is stable, and the sampling precision is high;
◆The positive and negative pressure combined sampling system can realize the sample degassing and is suitable for the test of different viscosity.
◆ Built-in air purification system to ensure the test is free from pollution;
◆ Built-in multiple calibration curve, the particle counter is compatible with all common standards at home and abroad for calibration;
◆ 8 common standards such as GJB-420B, NAS1638, ISO4406 and ГOCT17216, support customization;
◆Standard testing, and can set the required standards according to customer needs;
◆ A variety of sampling containers such as standard sampling bottles or sampling cups can be used to meet the testing requirements of different industries;
◆Color touch screen operation, built-in printer, simple and generous structure, easy to operate;
◆ Full-featured automatic operation, English input, with data storage and printing functions;
◆ Built-in data analysis system, which can determine the sample level according to the standard;
◆With RS232 interface, the particle counter can be connected to computer or laboratory platform for data processing;
| Applicable standard | GJB-420B, NAS1638, ISO4406 and ГOCT17216 etc |
| Light source | A semiconductor laser |
| Particle size range | 0.8um~500um |
| Detection channel | 8 channels set particle size arbitrary |
| Resolution | Above 10% |
| Repeatability | RSD<2% |
| Viscosity range | Maximum 350cSt |
| Sampling volume | 0.2~1000ml |
| Sample accuracy | 5mL/min~80mL/min |
| Maximum vacuum of Pneumatic chamber | 0.08MPa |
| Maximum positive pressure of Pneumatic chamber | 0.8MPa |
| Limit coincidence error | 10000 /mL |
| Power supply | AC220V±10%,50Hz |
| Dimension | 340*340*640mm |
| Net weight | 28KG |